User Guide Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product

User Guide Co-Dx ZDC-R-001 Logix Smart Zika 
Dengue and Chikungunya RUO Product
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product User Guide
Perform in Clean Area
1) Plate the Master Mix
a. Load the Co-Dx Box software.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Load the Co-Dx Box software
b. In the Clean area, clean work surfaces with water > 10% bleach > water > 70% alcohol (in that order), or another equivalent method of cleaning that disinfects and degrades nucleic acids.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - In the Clean area, clean work surfaces
c. Thaw the master mix (MM) and the nuclease-free water (NFW) (for the no-template control [NC]), on ice.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Thaw the master mix (MM)
d. Briefly vortex and centrifuge the MM and NFW reagents.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Briefly vortex and centrifuge the MM
e. Add 5 µL of MM into every well that will be used (1 well for NC, 1 well for positive control [PC], and 1 well for each sample).
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Add 5 µL of MM into every well that will be used
f. Pipet 5 µL of NFW into only the NC well and cap the NC well.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Pipet 5 µL of NFW into only the NC well and cap the NC well
g. Place reagents back into the freezer.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Place reagents back into the freezer
Perform in Positive Area
2 Plate Samples and Positive Control (PC)
a. Load the Co-Dx Box software.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Load the Co-Dx Box software 2
b. In the Positive area, clean work surfaces with water > 10% bleach > water > and 70% alcohol (in that order).
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - In the Positive area, clean work surfaces with water
C. Thaw the PC on ice.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Thaw the PC on ice.
d. Briefly vortex and centrifuge all reagents.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Briefly vortex and centrifuge all reagents.
e. Perform sample extraction, either using RNA extraction or lysis.
f. Add 5 µL of the sample extract to the appropriate wells and cap the sample extract wells.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Add 5 µL of the sample extract to the appropriate
g. Add 5 µL PC to the PC well and cap the PC well.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Add 5 µL PC to the PC well and cap the PC well
h. Place capped tubes in the thermocycler, in appropriate locations and fill the rest of the carousel with blank tubes.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Place capped tubes in the thermocycler
i. (Optional): Cap all tubes and place samples and reagents back into the freezer.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - (Optional) Cap all tubes and place sam- ples and reagents back into the freezer
Perform in Co-Dx Box Software
3 Start the Run
a. In the Co-Dx Box software, click New.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product -In the Co-Dx Box software, click New
b. Select an assay template and click OK. 
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Select an assay template and click OK.
c. Under Run Setup, select Samples and type sample information in the appropriate order.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Under Run Setup, select Samples and type sample information in the appropriate order
d. Click Information and in the Notes field,  type any needed information.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Click Information and in the Notes field, type any needed information
e. Select the appropriate thermocycler and from the drop-down select Start Run.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Select the appropriate thermocycler and from the drop-down select Start Run.
f. Verify the tube clamp is in place and the appropriate reaction volume is listed, and select Start.
Co-Dx ZDC-R-001 Logix Smart Zika Dengue and Chikungunya RUO Product - Verify the tube clamp is in place and the appropriate reaction volume is listed, and select Start
The run initializes and a progress bar gives an estimated time to completion.